Trio™ Operating Instructions
0070-10-0666-01
2 - 41
Operations
Parameter Menus
PR ALM HI
Allows the user to set the upper limit of the Pulse Rate alarm.
(See “Nellcor SpO2 Alarm Limits” on page 2-42.)
PR ALM LO
Allows the user to set the lower limit of the Pulse Rate alarm.
(See “Nellcor SpO2 Alarm Limits” on page 2-42.)
SWEEP
Adjusts the speed of the SpO
2
waveform on the display. The
selections are
12.5
and
25.0
mm/sec.
BEEP VOL
Beep volume is the volume of the audible pulse rate tone when SpO
2
is selected as the HR Source. The selections are
OFF
,
LOW
,
MED
and
HIGH
.
SENSOR OFF
AUDIO
Controls the onset of the audio beep alarm for the “SpO2 Sensor
OFF” condition. The selections are
ON
and
OFF
.
If ON is selected, then the audio beep alarm will sound when an
“SpO2 Sensor OFF” condition occurs. If OFF is selected, then the
audio beep alarm will not sound when an “SpO2 Sensor OFF”
condition occurs.
SAT SECONDS
Controls pulse oximetry nuisance alarms (see section 2.4.3.2 for
details on SatSeconds™ features). The selections are
OFF
,
10
,
25
,
50
or
100
.
When SatSeconds is enabled, the SpO
2
Low and High Alarm
functions are automatically disabled. As a safety precaution, if three
(3) SpO
2
alarm violations occur within 60 seconds, a priority 2 alarm
will trigger even if the SatSeconds limit has not been reached.
RESTORE
DEFAULTS
Allows the user to restore the SpO
2
user default configuration.