Parameter Menus
Operations
2 - 40
0070-10-0666-01
Trio™ Operating Instructions
CAUTION:
2.4.3.2.1
Nellcor
®
SpO
2
Setup Menu
FIGURE 2-34
Nellcor SpO
2
Setup Menu
Accessing the Nellcor SpO
2
Setup Menu
To access the
NELLCOR SPO2 SETUP
menu, select
SpO2
from the
SpO2
parameter tile
using the Navigator™ Knob. Once in the
NELLCOR SPO2 SETUP
menu, use the
Navigator Knob to adjust settings. To close the menu, select
NORMAL SCREEN
(from the
menu or the Front Panel Keypad).
Nellcor SpO
2
Setup Menu Selections
ALM PRIORITY
Enables the user to select the priority of the SpO
2
alarm. Choices are:
1
,
2
and
3
. Priority 1 alarms are considered the most serious.
ALM PRINT
Enables or disables automatic printing during an SpO
2
alarm
condition. Choose
ON
to enable printing upon SpO
2
alarm. Choose
OFF
to disable printing upon SpO
2
alarm.
SPO2 ALM HI
Allows the user to set the upper limit of the SpO
2
alarm.
(See “Nellcor SpO2 Alarm Limits” on page 2-42.)
SPO2 ALM LO
Allows the user to set the lower limit of the SpO
2
alarm.