B-4 Electrical Safety Inspection
normal polarity( Normal Condition);
reverse polarity( Normal Condition),
normal polarity with open neutral(Single Fault Condition);
reverse polarity with open neutral(Single Fault Condition).
normal polarity with open earth(Single Fault Condition);
reverse polarity with open earth(Single Fault Condition).
LIMITS
For CF
applied parts
10
μ
A in Normal Condition
50
μ
A in Single Fault Condition
For BF
applied parts
100
μ
A in Normal Condition
500
μ
A in Single Fault Condition
B.8
Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage,
through a limiting resistance, to selected applied part terminals. Current measurements are
then taken between the selected applied part and earth. Measurements are taken with the
test voltage (110% of mains) to applied parts in the normal and reverse polarity conditions
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal
Polarity;
Reversed
Polarity
LIMITS
For
CF applied parts: 50
μ
A
For
BF applied parts: 5000
μ
A
B.9 Patient
Auxiliary
Current
Patient Auxiliary currents are measured between any selected Applied Part connector and the
remaining Applied Part connector. All measurements may have a true RMS only response.
The following outlet conditions apply when performing the Patient Auxiliary Current test.