Probes and Biopsy 13-19
NGB-011
13.2.3 Needle-guided Bracket Inspection and
Installation
Inspection of the Needle-guided Bracket
Be sure to perform inspections before and after use of the needle-guided bracket. If an
abnormality is found on the needle-guided bracket, immediately stop using it and contact
MINDRAY Customer Service Department or sales representative.
1. Sterilize the needle-guided bracket before and after use.
2. Confirm that the needle-guided bracket is free of damage, deformation, stripping,
and malfunction, loose or missing parts.
3. Confirm that the needle-guided bracket is securely mounted in the correct position.
Installing the Needle-guided Bracket
NGB-004
(1) Put on the sterile probe sheath.
Needle guide rack
Needle guide
hole
Grip
knob
Locating groove
Clamp
Clamping knob of the
needle guide
Needle guide
Locating
pit
Probe
Needle-guided
bracket