EMC Guidance and Manufacturer’s Declaration 15-3
TABLE 2
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The customer
or the user of system should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT-GUIDANCE
Electrostatic
Discharge(ESD)
IEC 61000-4-2
±6 kV contact;
±8 kV air
±6 kV contact;
±8 kV air
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
Transient / burst
IEC 61000-4-4
±2 kV for power
supply lines;
±2 kV for power
supply lines;
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s);
±2 kV line(s) to
earth
±1 kV line(s) to
line(s);
±2 kV line(s) to
earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, Short
interruptions and
voltage variation on
power supply input
voltage
IEC 61000-4-11
<5% U
T
(>95%
dip in U
T
) for 0.5
cycle
40% U
T
(60% dip
in U
T
) for 5 cycle
70% U
T
(30% dip
in U
T
) for 25 cycle
<5% U
T
(>95%
dip in U
T
) for 5
sec
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycle
70% U
T
(30% dip in
U
T
) for 25 cycle
<5% U
T
(>95% dip
in U
T
) for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If you
require continued operation
during power mains
interruptions, it is recommended
that our product be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 HZ) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE:
U
T
is the A.C. mains voltage prior to application of the test level.