Power Network Application Trainer
– Earthing / Grounding network impedance measurement
89
4.1.1.3 Measurement with MI 3295S Step Contact Voltage Measuring
System
Figure 63-2: Earth measurement with MI 3295S
This exercise will present the usage of MI 3295S measuring system for earth resistance
measurement. Four wire earth resistance measurement method is supported.
Measurement is done at fix frequency 55 Hz. MI 3295S device needs external 230V AC
power supply and it is intended for measurement of bigger systems, where higher test
current needs to be generated. Test current up to 7.5 A could be generated and test
voltage up to 50 V.
Switch position Simulated condition
Application
Expected result
Regular measurement place
condition (simulated resistive earth
character), single and sweep
frequency measurement
Radio towers
Pylons without ground wire
connection
Solar power plants
Wind & water turbine
Power transmission
Power distribution
Re = 2,94
Ω @ 55 Hz
Table 17-2: Measurement setup/result for 4-wire measurement method with MI 3295S
Measuring procedure:
1. Select proper switch position on the training module.
2. By using function selection keys (Left & Right) select
EARTH
measurement function
and with navigation buttons (Up & Down) select
EARTH RE
function.
Summary of Contents for MI 3298
Page 55: ...Power Network Application Trainer Earthing Grounding network impedance measurement 55 ...
Page 56: ...Power Network Application Trainer Earthing Grounding network impedance measurement 56 ...
Page 124: ...Power Network Application Trainer Earthing Grounding network impedance measurement 124 ...
Page 228: ...MI 3290 Earth Analyser 228 ...
Page 229: ...MI 3280 Digital Transformer Analyser 229 ...
Page 230: ...MI 3280 Digital Transformer Analyser 230 ...
Page 231: ...MI 3250 MicroOhm 10A 231 ...
Page 232: ...MI 3205 TeraOhmXA 5kV 232 ...