38
Solar Blue User Manual
Chapter 2
LEDs and button wireless puller
The following table describes what the LED colors and status indicate:
LED
Color
Indication
Bluetooth
Blue blinking
The Bluetooth catheter puller has a wireless
connection with the computer.
Status
Green on
The puller is switched on and is functioning
well.
Orange on
IPL mode and/or downloading firmware.
Red on
The puller has detected an internal failure.
Contact your LABORIE representative.
Battery
Green on
The rechargeable batteries are fully charged.
Green blinking
The power adapter is connected to the puller
(and charging the batteries).
Orange blinking
The batteries are almost empty, please connect
the power adapter (within five minutes).
Red blinking
The batteries are depleted. You can connect
the power adapter to perform the investigation
and to charge the batteries. If you don’t follow
up on the instruction, the puller will be switched
off and lose the wireless connection with the
computer.
During the pre-test and the actual investigation, the slider is locked in position.
To unlock and reposition the slider manually, you must press the