Anorectal Manometry
Solar Blue User Manual
105
Figure 9.11
Sensation settings
During the sensation test, different sensation markers can be placed. The
following markers can be enabled in the protocol: no sensation, first sensation,
first urge, modest urge, intense urge, maximum tolerated volume, and pain.
The setting
Display balloon markers
can be (de)selected.
Balloon filling settings
On the
Filling
tab, you can set the filling method. For the
Syringe
filling
method, you can enter the volume increment (e.g., 10 ml) or edit a complete
list (e.g., 10-10-10-20-20). You can also select
Ask balloon volume
. After
inserting a sensation marker during the test, the software will ask to enter the
balloon volume.
For filling the balloon with the pump (water) you can enter the pump speed
(default 50 ml/min) on the
Filling
tab.