Explanation of symbols
Solar Blue User Manual
237
Message
Cause
Posible actions to
resolve
The system is not powered. Please check the
power connection
System is not powered
Connect the power
The WPU [Name] has lost the connection
Communication problems
Contact your LABORIE
representative
There are more channels available in the
protocol file ([Filename]) then supported by this
software version!
Protocol file comes from a later
version
Install correct software
version or contact your
representative
This color scheme is read-only
Operator tries to change a system
color scheme
Modify a different color
scheme
Unable to display the Clouse contour plot
Problem displaying graphs
Contact your LABORIE
representative
Unable to display the graphs
Problem displaying graphs
Contact your LABORIE
representative
Unable to start the investigation (MMS
perfusion pressure too high) The pressure
release valve is probably polluted. Release the
tube from the water container and try to restart
the pump. If the error remains, the system
should be sent to MMS for maintenance.
Perfusion problem
Contact your LABORIE
representative
Warning: compliance marker in miction phase
Compliance marker in miction
phase
Remove marker
Warning: No more Profile markers!
No more profile markers available Remove profile marker
Warning: No more Profile markers!
No more profile markers available Remove profile marker
Wireless patient module connection lost.
Communication problems
Contact your LABORIE
representative
WPM not connected! Check the Range of the
Wireless patient module, Check the Bluetooth
dongle or replace the batteries
Module expected but not
connected
Contact your LABORIE
representative
WPM not connected! Check the Range of the
Wireless patient module, Check the Bluetooth
dongle or replace the batteries
Module expected but not
connected
Contact your LABORIE
representative
WPU not connected! Check the Range of the
Wireless patient unit, Check the Bluetooth
dongle or replace the batteries
Module expected but not
connected
Contact your LABORIE
representative
WPU not connected! Check the Range of the
Wireless patient unit, Check the Bluetooth
dongle or replace the batteries
Module expected but not
connected
Contact your LABORIE
representative
Writing of channel [Channel name] to catheter
file failed!
Writing to catheter file failed
Retry calibration
Software messages database
Message
Cause
Posible actions to
resolve
Can not execute [application]
Application cannot be executed Contact your MMS
representative
Chipcard error
Chipcard could not be read
Contact your MMS
representative
Could not open combination report because
of a problem with following investigation:
[Report file] Generate report file by
N/A
Contact your LABORIE
representative