100
Solar Blue User Manual
Chapter 9
Figure 9.5
Filling with syringe settings
The option
Ask balloon volume
is available for the syringe filling method
(most common method) and can be used for the push test, RAIR test,
sensation test, and balloon expulsion test. After inserting the end marker
during the test, the software will ask to enter the balloon volume.
The
Volume increment
setting and the
List
selection is applicable for the
syringe filling method for the Sensation test. You can enter the volume
increment (e.g., 10 ml) or edit a complete list (e.g., 10-10-10-20-20).
9.2.5
Timers
On the
Timers
tab (Figure 9.6), you can enable timers and set the duration.
Timers will help you to perform the test with a certain duration and to show this
to the patient. During the test, a progress bar will appear on the screen as
soon as the start of test has been marked. When the time elapses, the
progress bar disappears, and the end marker is automatically placed.