56
Solar Blue User Manual.
Chapter 3
3.5.5
Flushing and zeroing water catheters
Water catheters and DT-XX pressure transducers must be flushed and zeroed
(Figure 3.16) before starting the investigation.
Figure 3.16
Flush and zero with syringe
The procedure is as follows:
▪
Catheters and transducers have been prepared as described in the
previous subsections; the transducer holder is at symphysis pubis level.
▪
Connect the syringes filled with sterile water to the pressure transducers as
displayed in the figure above (optionally, the LABORIE flush set can be
used).
▪
Push the blue flushing devices and flush the pressure transducers and
catheter lumens to expel all air bubbles.
▪
Release the flushing devices to stop flushing.
▪
Inspect the pressure transducers, catheter lumens and flow-reducer
visually for air bubbles. If necessary, you must flush them again.
▪
Turn all stopcock handles counterclockwise, the text “OFF” pointed to the
catheter side.