Anorectal Manometry
Solar Blue User Manual
123
displayed on the
Pump
button in the upper-left corner of the screen
(e.g., 10 ml).
Press the
Pump
button to start the pump. The balloon in the rectum
will be slowly filled.
When the desired volume is reached, stop the pump by pressing the
Pump
button.
Press the
Balloon expulsion
button to mark the start of the test. A
balloon marker is automatically inserted. Ask the patient to expel the
balloon.
Press the
Balloon expulsion End
button to mark the end of the test. A
dialog box appears (Figure 9.20), asking whether the balloon expulsion
was successful. Select
Failed
,
Partial
, or
Success
and then click
OK
.
9.4.12
Rectal compliance
The purpose of the compliance measurement is to objectively determine the
compliance of the rectum. The compliance can be used to determine
abnormalities in the elasticity of the rectum.
Balloon compliance correction
To obtain the real compliance of the rectum, it is important to determine the
compliance of the balloon itself. To calibrate the balloon compliance, proceed
as follows. Place the balloon in warm water at body temperature (37°C). Make
sure that the following configurations are set in the protocol (see § 9.2.10):
•
On the Filling tab:
Filling method: Syringe
Volume increment: 10 ml
•
On the Sensation/Compliance tab:
Use std. Balloon calibration
Always
Save std. Balloon calibration
Always
Click the
Anorectal manometry
button to start the pre-test.