Safety information
Solar Blue User Manual
215
WARNING
Do not connect an additional Multiple Socket-Outlet
(MSO) or extension cord to the Solar Blue system.
WARNING
Do not connect non-ME equipment that is supplied
as part of this system intended to be plugged into
the Safety Module into a wall outlet. The
Safety module is a safety component designed to
ensure Touch/Enclosure Leakage currents are
below the allowable limits. Operator may
experience electrical shock if non-ME equipment is
connected directly to the wall outlet.
WARNING
Only connect equipment that has been supplied as
part of the Solar Blue system to the Multiple Socket
Outlet provided on the Separating transformer.
CAUTION
Do not position the Solar Blue system so that it is
difficult to operate the disconnection device.
Installation
WARNING
Safety can be violated if the Solar Blue measurement
system or one of its modules shows visible damage.
WARNING
Do not place any equipment in front of the mains plug; to
isolate the system, the mains plug must remain easily
accessible.
Solar pole (Mk III and Mk IV)
WARNING
If a safe system setup, in accordance with IEC-60601
cannot be guaranteed, a safety module is required.
Please contact your LABORIE representative.
WARNING
Without a properly installed safety module, the laptop’s
power must be supplied by its batteries only. The laptop
must also not be connected to a printer or any other