Chapter 5 Disassembly and Assembly
5-119
CCU INFO screen
•
Host name: R170-0001
•
Host name (on JobManager): r170-0001
•
IP address: 192.168.20.170
PostgreSQL access tool
•
c_status: set to "CS#-0001" only
•
r_status: set to "r170-0001" only
•
relations: set to a single relation between "CS#-0001" and "r170-0001" only
•
sys_config: set to "1"
(CS#: "CS1" for CS-1 or CS-3; "CS2" for CS2)
Alternatively you can also initialize the JM settings using a service tool.
4
Exit the REGIUS console and the equipment and turn each power off.
5
Replace the CF card.
1. See
Removal Procedures (Page 5-6)"
"5.2.3 Removing/Installing the Exterior Panel and
to remove the first front back panel.
2. Remove the circuit board cover.
•
3 screws (M3 x 6)
Loosen all.
3. Replace the CF card.
4. Install the circuit board cover.
•
3 screws (M3 x 6)
5. See
Installation Procedures (Page 5-8)"
"5.2.3 Removing/Installing the Exterior Panel and
to install the first front back panel.
First Front Back Side
CF
Card
First Front Back Side