Chapter 4 Confirming Operation Using Service Tool
4-2
4.1
Before Confirming the Operation
4.1.1 Outline of Confirming the Operation
To confirm the operation of each unit of the REGIUS MODEL 110 or confirm the status of the sensors,
operate the [Unit Test] screen in the service tool (only for the REGIUS MODEL 110) on the REGIUS
console.
The equipment will go into maintenance mode once the [Unit Test] screen is displayed, and only the
commands from the REGIUS console displaying the [Unit Test] will be accepted (same status as
performing the calibration).
Following operation can be performed on the [Unit Test] screen:
•
Stand-alone operation of each unit
•
Step operation (perform each step in order of read sequence, and display OK/NG of each operation)
•
Display status of the sensors
•
Setup of the reader parameter such as volume
•
Setup of compensation value such as justifier (operation for the setup is possible too)
Network Hub
REGIUS
Console
REGIUS
MODEL 110
[Unit Test] screen