Chapter 3 Troubleshooting
3-40
3.4.2 Image is Dark or Black in General (Still With Structure of the Object)
3.4.3 Image is Bright or White in General (Still With Structure of the
Object)
No.
Cause
Response
1
Output density in the G process
parameter is set high in general.
See the Image Process Adjustment Manual and set
the Density H/L in the G process parameter lower.
2
X-ray dosage is too much (in case of fix
the process or unprocessed image).
Confirm the exposure condition if the X-ray dosage is
not too much.
3
S value (gain) set in the REGIUS console
is too large (in case of fix the process or
unprocessed image).
Set the S value to the appropriate value.
4
Sensitivity calibration problem (in case of
fix the process and unprocessed image).
Redo the sensitivity calibration.
5
The ROI recognized for process was not
setup correctly for some reason.
Set the ROI to the correct position.
No.
Cause
Response
1
Output density in the G process
parameter is set low in general.
See the Image Process Adjustment Manual and set
the Density H/L in the G process parameter higher.
2
X-ray dosage is extremely low or not
generated at all.
•
Confirm the exposure condition.Confirm that the
exposure switch is not released during the X-ray
exposure.
•
Confirm the X-ray dosage using the dosimeter.
3
S value (gain) set in the REGIUS
console is too small (in case of fix the
process or unprocessed image).
Set the S value to the appropriate value.
4
The ROI recognized for process was not
setup correctly for some reason.
Set the ROI to the correct position.
5
Photomultiplier switch on the LMC is
turned OFF.
Turn it ON.
6
Problem with the LMC.
Replace the LMC.
See
"5.6.2 Replacing the LMC (Page 5-79)"
for the
procedure to replace the LMC.
7
Failure of the photomultiplier.
Replace the photo multiplier filter assy.
See
"5.6.4 Replacing the Photomultiplier Tube Filter
for the procedure to replace the
photomultiplier filter or assy.
Summary of Contents for 0902
Page 2: ......
Page 6: ......
Page 12: ......
Page 30: ...2 10...
Page 54: ......
Page 84: ...Chapter 3 Troubleshooting 3 30 3 2 19 Confirm the H sync Signal To be stated...
Page 85: ...Chapter 3 Troubleshooting 3 31 3 2 20 Confirm the LMC Operation To be stated...
Page 120: ......
Page 296: ...8 10...
Page 300: ......
Page 301: ......