Foot Pedal Setup
Z370584 Rev. D
5-20
Surgeon Settings
7. Tap the
Finished
button to:
accept your choice of switch assignment
close the Switch Assignment dialog box
return to the edit screen
Repeat the process for all switches.
8. Tap the
Save
button to retain the switch assignments.
9. Tap
Exit
to close the Foot Pedal Set Up screen.
VERITAS Foot Pedal Threshold Configuration
Follow the steps below to configure the thresholds for the foot pedal.
NOTE: In addition to showing the type and status of the foot pedal, the foot pedal icon
(shown below) in the top of the screen also allows access to the Foot Pedal Settings
screen.
1.
On the Surgeons and Programs screen, tap
or
.
2. Tap the
Thresholds
button under Foot Pedal. The system displays the foot pedal
thresholds screen.
Bottle Down
Lowers the IV pole while the switch is held down.
Reflux
Activates reflux.
Continuous Irrigation
Activates or deactivates continuous irrigation.
1-Touch Up
Increments CASE mode 1-Touch. (Phaco only)
1-Touch Down
Decrements CASE mode 1-Touch. (Phaco only)
Toggle CASE
Activates or deactivates CASE MODE (Phaco only)
Switch Setting
Description
Summary of Contents for Veritas
Page 2: ...Z370584 Rev D This page was intentionally left blank...
Page 4: ...Z370584 Rev D This page was intentionally left blank...
Page 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Page 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Page 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Page 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Page 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Page 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Page 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Page 270: ...Z370584 Rev D I 8 This page was intentionally left blank...