Surgeon Settings
Z370584 Rev. D
Foot Pedal Setup
5-13
Figure 5-11 — Single Linear Foot Pedal Thresholds
4. Check the
Feedback
checkbox to enable foot pedal feedback. Feedback is a physical
and audible click as you move the foot pedal to the next or previous position.
5. Tap
Save
to retain the settings.
Single Linear Foot Pedal Calibration and Testing
Follow these steps to calibrate and test your foot pedal.
1.
On the Surgeons and Programs screen, tap
.
2. Tap the
Calibration
button.
NOTE: The Calibration feature is not available during surgery.