Foot Pedal Setup
Z370584 Rev. D
5-14
Surgeon Settings
Figure 5-12 — Single Linear Foot Pedal Calibration Screen
3. With your foot off the foot pedal, tap the
Phase 1
button.
4. Put your foot on the foot pedal, press and hold the foot pedal all the way down and tap
the Phase 2 button.
5. The system automatically displays the Foot Pedal Test screen.
6. Press the foot pedal down fully to position 3. As the foot pedal travels through positions 1
and 2 to position 3, the progress bar shows its current state. When the pedal reaches the
bottom of position 3, the progress bar should be solid green.
7. Press each of the four switches. When you press a switch, the On button for that switch
turns green and its Off button turns white. When you release the switch, the On button
for that switch turns white and its Off button turns green.
8. Tap the
Finished
button when you have completed testing the foot switches.