Status, Warning, and Error Messages
Z370584 Rev. D
10-24
Error Messages, Troubleshooting, and Diagnostics
1.
Select
End Case
.
2. Select
Next Case
.
3. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
311 Dump valve pressure high/low.
Probable cause:
•
Piston pump not working.
•
The VIT dump valve not closing.
Corrective action:
1.
Select End Case.
2. Select Next Case.
3. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
327 Pneumatics system pressure too low.
Probable cause:
•
The system pressure is too low.
•
The piston pump does not work.
•
There is a system air leak.
•
The vitrectomy cut valve is on.
Corrective action:
1.
Select
End Case
.
2. Select
Next Case
.
3. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
328 Pneumatics system pressure too high.
Probable cause:
•
The system pressure is too high.
•
The external air valve does not work.
•
The piston pump does not work.
•
The selector valve does not work.
Corrective action:
1.
Select
End Case
.
Summary of Contents for Veritas
Page 2: ...Z370584 Rev D This page was intentionally left blank...
Page 4: ...Z370584 Rev D This page was intentionally left blank...
Page 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Page 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Page 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Page 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Page 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Page 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Page 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Page 270: ...Z370584 Rev D I 8 This page was intentionally left blank...