Vitrectomy Specifications
Z370584 Rev. D
12-20
Specifications
Vitrectomy Specifications
Table 12.17 – Vitrectomy Specifications
Diathermy Specifications
Table 12.18 – Diathermy Specifications
Peristaltic Pump
Vacuum Level
0–650 mmHg in 5 mmHg increments
Pump Flow Rate
0-60 cc/minute, in 1 cc increments
Venturi Pump
Vacuum Level
0-600 mmHg in 5 mmHg increments
Atmospheric
Pressure/Altitude
Density Altitude (DA) conditions are 2,500 cpm from sea level to 3,000 feet
(914.4 m) elevation and shall be de-rated by 100 cpm per 1,000 ft (304.8 m)
additional elevation up to a maximum of 10,000 ft (3,048 m) above sea level.
Indicated Cut Rate
Actual Cut Rate
Upper Limit (cpm)
Actual Cut Rate Lower Limit (cpm)
2500
3000
2000
Power Adjustment
5% to 100%, in 5% increments
Diathermy Power (maximum power into rated load)
8.5 Watts into 350 Ohms 386 kHz
Maximum Output Current
0.17 Amps
Diathermy Type
Bipolar
Rated Peak Voltage
160 Volts
Power Cord Length
12 ft (3.6 m)
Connector of diathermy forcep/probe
Ø.072 RECEPTACLE (2)
All measurements are in inches.
Summary of Contents for Veritas
Page 2: ...Z370584 Rev D This page was intentionally left blank...
Page 4: ...Z370584 Rev D This page was intentionally left blank...
Page 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Page 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Page 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Page 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Page 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Page 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Page 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Page 270: ...Z370584 Rev D I 8 This page was intentionally left blank...