Status, Warning, and Error Messages
Z370584 Rev. D
10-16
Error Messages, Troubleshooting, and Diagnostics
149 Fluidics strain gauge error.
Probable cause:
•
The Fluidics Tubing pack is not properly mounted.
•
Bad strain gauge.
Corrective action:
1.
Reinsert the Fluidics Tubing pack.
2. Prime the system.
3. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
151 Peristaltic pump stall.
Probable Cause: Encoder, Decoder, or Stepper driver is faulty.
Corrective Action:
1.
Reinsert the Fluidics Tubing pack.
2. Prime the system.
3. If the error does not clear or if the error continues to occur, document the error message
and contact JJSV for technical service.
160 Advanced Infusion pressure is too low.
Probable Cause: Pressure sensor malfunction, regulator malfunction, leaks at connection to
console or spike.
Corrective Action:
1.
If changing the balanced salt solution, proceed and error will clear when spike is
reinserted.
2. Otherwise, check for leakage or loose connector at the Fluidics Tubing pack and console
and re-prime.
3. If error continues, try changing the Fluidics Tubing pack and re-prime.
4. If the error continues to occur, use an Advanced Fluidics Program to continue and
contact JJSV for technical service.
161 Advanced Infusion pressure is too high.
Probable Cause: Pressure sensor or pressure regulator is faulty.
Corrective Action:
1.
Change to an Advanced Fluidics Program.
2. Disconnect the infusion connector from the console if connected.
3. Prime the system to continue.
4. If the error continues to occur, contact JJSV for technical service.
Summary of Contents for Veritas
Page 2: ...Z370584 Rev D This page was intentionally left blank...
Page 4: ...Z370584 Rev D This page was intentionally left blank...
Page 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Page 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Page 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Page 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Page 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Page 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Page 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Page 270: ...Z370584 Rev D I 8 This page was intentionally left blank...