Z370584 Rev. D
6-28
Program Settings
Figure 6-18 — Phaco Surgical Screen with Occlusion and CASE On
Occlusion Mode Settings
With Occlusion Mode, you can set different aspiration rates for occluded aspiration as
opposed to unoccluded aspiration.
You can set a different vacuum rise time for when the phaco tip occludes without changing
the aspiration rate through an unoccluded tip.
To set the occluded aspiration rate thresholds:
1.
Tap
Phaco
.
2. Tap
Settings
.
3. Check the
Occlusion
checkbox to enable Occlusion Mode.