Program Settings
Z370584 Rev. D
Program Settings
6-9
Irrigation and Aspiration (I/A)
1.
Tap
to open the Settings screen.
2. Select an Irrigation/Aspiration submode on the left (
Figure 6-7 — Irrigation/Aspiration Settings
3. If desired, tap the submode name at the top of the screen (outlined in red in
customize the submode name.
4. Select the pump type (
Peristaltic
or
Venturi
).
5. Use the up and down arrows to set the
Aspiration
rate (0 to 60).
6. Use the up and down arrows to set the
Vacuum
. If you are using Venturi, the range is 0 to
600. If you are using Peristaltic, the range is 0 to 650.
7. Select the delivery mode (
Linear
or
Panel
).
8. Set the
Bottle Height
(Advanced Fluidics) or
Total Height
(Advanced Infusion).
9. Tap
Save
to save the program.