Specifications
Z370584 Rev. D
Phacoemulsification Specifications
12-19
Table 12.14 – Wireless Quality of Service and Security
Phacoemulsification Specifications
Table 12.15 – Phacoemulsification Specifications
Irrigation and Aspiration Specifications
Table 12.16 – Irrigation and Aspiration Specifications
Device
RF Wireless
Technology
Part Number
Operating
Frequencies
Operating
Distances
Industry
Standard
Quality of
Service and
Security
VERITAS
Vision
Console
Bluetooth
Texas
Instruments
CC2541
2.402 to
2.480 GHz
Up to 3
meters
Bluetooth
Specification
version 4.0
FCC CFR47
Part 15,
Subpart C
“Intentional
Radiators”
The VERITAS
Vision Console
wireless
communication
has low latency,
high data
integrity, and is
robust in the
presence of
interference as
defined by the
Bluetooth
specification.
Phaco Power
1%-5% in 1% increments,
5-100%, in 5% increments
Peristaltic Pump
Vacuum Level
10–650 mmHg in 5 mmHg increments
Pump Flow Rate
10-60 cc/minute
Venturi Pump
Vacuum Level
0-600 mmHg in 5 mmHg increments
Peristaltic Pump
Vacuum Level
0–650 mmHg in 5 mmHg increments
Pump Flow
0-60 cc/minute
Venturi Pump
Vacuum Level
0-600 mmHg in 5 mmHg increments
Summary of Contents for Veritas
Page 2: ...Z370584 Rev D This page was intentionally left blank...
Page 4: ...Z370584 Rev D This page was intentionally left blank...
Page 22: ...Z370584 Rev D 1 12 This page was intentionally left blank...
Page 36: ...Z370584 Rev D 2 14 This page was intentionally left blank...
Page 54: ...Z370584 Rev D 3 18 This page was intentionally left blank...
Page 122: ...Z370584 Rev D 5 42 This page was intentionally left blank...
Page 158: ...Z370584 Rev D 6 36 This page was intentionally left blank...
Page 184: ...Z370584 Rev D 7 26 This page was intentionally left blank...
Page 198: ...Z370584 Rev D 9 4 This page was intentionally left blank...
Page 270: ...Z370584 Rev D I 8 This page was intentionally left blank...