Power Supply Specification Guidelines
Intel® Server Board S2600CW Family TPS
146
Revision 2.4
AC HOT
POWER SUPPLY
AC NEUTRAL
V
OUT
RETURN
V
AC GROUND
LOAD
SCOPE
LOAD MUST BE
ISOLATED FROM
THE GROUND OF
THE POWER
SUPPLY
10uF
.1uF
GENERAL NOTES:
1. LOAD THE OUTPUT WITH ITS MINIMUM
LOAD CURRENT.
2. CONNECT THE PROBES AS SHOWN.
3. REPEAT THE MEASUREMENTS WITH THE
MAXIMUM LOAD ON THE OUTPUT.
SCOPE NOTE:
USE A TEKTRONIX 7834 OSCILLOSCOPE WITH 7A13 AND
DIFFERENTIAL PROBE P6055 OR EQUIVALENT.
Note
: When performing this test, the probe clips and capacitors should be located close to the load.
Figure 32. Differential Noise Test Setup
10.3.2.15
Timing Requirements
Below are timing requirements for the power on/off of the PDB DC/DC converters. The +3.3V,
+5V and +12V output voltages should start to rise approximately at the same time. All outputs
must rise monotonically.
Table 83. Output Voltage Timing
Description
Min
Max
Units
Output voltage rise time for each main output; 3.3V, 5V, -
12V, and 5Vstby.
1.0
20
msec
The main DC/DC converters (3.3V, 5V, -12V) shall be in
regulation limits within this time after the 12V input has
reached 11.4V.
20
msec
The main DC/DC converters (3.3V, 5V, -12V) must drop
below regulation limits within this time after the 12V input
has dropped below 11.4V.
20
msec
The 5Vstby converter shall be in regulation limits within this
time after the 12Vstby has reach 11.4V.
20
msec
The 5Vstby converter must power off within this time after
the 12Vstby input has dropped below 11.4V.
20
msec
10.3.2.16
Residual Voltage Immunity in Standby Mode
Each DC/DC converter is immune to any residual voltage placed on its respective output
(typically a leakage voltage through the system from standby output) up to 500mV. This
residual voltage does not have any adverse effect on each DC/DC converter, such as: no
additional power dissipation or over-stressing/over-heating any internal components or
adversely affecting the turn-on performance (no protection circuits tripping during turn on).
While in Stand-by mode, at no load condition, the residual voltage on each DC/DC converter
output does not exceed 100mV.