Note
18. Take full screen display on the scope. Check that the transition
times <4.8 ns in both normal and complement modes.
19. Set the HP 8 1 12A to Linear transition.
20. Measure the transition times are <6.2 ns.
Transition times can be increased by increasing the overshoot. If
Overshoot/Transition times adjustment cannot be achieved within
specification, change values of A1C532/C409 (See table 9- 1) and
repeat steps 1 to 15.
Adjustment Procedures 9-9