Store and Recall
Function Test
6. Set up the HP 81 16A Pulse/Function Generator as follows:
Output
Squarewave
Frequency
1 kHz
Amplitude
2.0 V
Offset
0.00 V
7. Verify that the scope display indicates a settling time of less than
200 µs as shown below.
'
Set t l i n g t i m e
.�
'
Figure
8-21 .
Correct Delay Control
100%
95%
Characteristics
Nine programmable locations for user preferred mode and parameter
settings
One for the standard (switch on) instrument mode and parameter set
(RCL 0).
One non accessible location for currently active mode and parameter
settings.
P�o,��ure
•
Press
•
Use right hand
VERNIER
key to obtain display
no
•
Press centre
VERNIER
key to recall the standard parameters.
•
Alter one of the parameters.
•
Press
•
Use right hand
VERNIER
key to obtain display UX, where X is
any number from 1 to 9.
•
Press left hand
VERNIER
key to store the altered parameter.
Testing Performance
8-29
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......