Test Equipment
Test Preparati,pn
Power Supplies
Refer to Chapter 1
Introduction
for the recommended test equipment
and accessories.
Refer to Chapter 10.1
Troubleshooting
for instructions on how to
open up the HP 81 12A instrument and prepare for servicing>
Control Board
�
M icroprocessor Board
0
0
0 0
Figure 9-1. Access to the HP 8 1 1 2A for Adjustments
Equipment
Digital Voltmeter (HP 3456A)
Procedure
1 . Connect the DVM low terminal to the ground testpoint on board
9-4 Adjustment Procedures
A l .
2. Test the supply voltages and, i f necessary, make adjustments to
achieve the levels given here:
Testpoint Adjust
Result
Al+l5 V A1R24 +15.000 V ±15 mV
Al-5.4 V A1R12 -5.40 V ±10 mV
A1+5.0 V
-
+5.050 V ±50 m V
A3+5.0 V
-
+5. 150 V ±50 mV
A 1+23 V A1R18 +23 .000 V ±50 mV
Al -23 V A1R19 -23.000 V ±50 m V
A l - 1 5 V A1R25 - 1 5.000 V ±15 mV
3. Disconnect the DVM.
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......