Procedure
1 . Connect the equipment as shown in the setup figure
2. Set up the HP 8 1 12A as follows:
Delay Control
Verification Test
Trigger Mode
Control Mode
Transition
DEL
HIL
101
NORM
PERC
Fixed
65 ns
2.00 v
0.00 v
3. Vary the power supply
(
or HP 3324 in DC Mode
)
between
approximately 1 volt and 10 volts, and verify that the counter
display indicates a range of periodic times which agree with the
specified limits for each of the HP 8 1 1 2A settings below.
HP 8112A
Counter
setting
reading
Low
High
100 ns
10 ns
20 ns
100 ns
10 µs
1
µs
1
µs
10
µs
10 ms
1 ms
1 ms
lO ms
1000 ms 100 ms 100 ms 1 .0 s
Characteristics
Pulse Delay ratio
1 :10
Control Voltage
1 .0 V to 1 0 V
Delay Ranges
Bandwidth
1 0 ns to 1 .0 s in eight non-overlapping decade
ranges. The fixed 55 ns delay of the instrument
has to be added to the delay induced by the
CTRL voltage
1 kHz
Equipment
•
Oscilloscope
(
HP 54121 T
)
•
Counter
(
HP 5335A
)
•
Variable DC Source
(
HP 6237B or HP 3324A)
•
Cable Assembly BNC
(
7 off
)
•
Attenuator 20 dB, 2 W
(
2 x )
•
BNC to Banana plug adaptor
•
BNC TEE piece
8-24
Testing Performance
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......