Specifications
Introduction
Timing Parameters
Common Specifications
All specifications in the following sections describe the warranted
performance of the instrument:
•
Timing parameters
•
Output parameters
•
Waveform characteristics
2
All specifications apply with a 50 n load, after a 30 minute warm-up
period, and are valid for ambient temperature in the range 1 5°C to
35°C. Refer to the General Characteristics section of this chapter for
the performance derating factor to be used outside this temperature
range
(
within the specified operating range of 0°C to 55°C
)
.
All operating characteristics given in the following sections describe
typical performance figures which are non-warranted:
•
Trigger modes
•
Control modes
•
Output Modes
•
Inputs and Outputs
•
Additional features
•
General characteristics
Unless otherwise stated, specifications are quoted for 503 amplitude
in normal mode, fastest transitions.
Resolution
3 digits, best case 100 ps
Accuracy
± 53 of programmed value ±2 ns
Repeatability
Factor 4 better than accuracy
Jitter
max 0.2% of programmed value + 100 ps
Specifications 2-1
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......