Output Levels
Performance Test
Specifications (values
in parenthesis into open
High Level Range
circuit)
- 7.90 V to 8 .00 V ( - 15.8 V to 16.0 V)
Low Level Range
-8.00 V to 7.90 V ( - 16.0 V to 15.8 V)
Level Accuracy
±40 mV ± 1 % of programmed value ±3% of amplitude
Repeatability
Factor 4 better than accuracy
Equipment
•
Digital Voltmeter (HP 3458A)
Test Setup
•
Two Cable Assemblies BNC (same length)
•
son
feed through connector ( ±0. 1 % )
HP 8112A
0
0
50
Ohm Feedthrough connector
and BNC to dual banana plug adapter
Voltmeter
(HP 3458A)
. .
.
.
Connected to
EXT TRIGGER
on rear of DVM
Figure
8-8.
Output Levels Performance Test
Procedure
1 . Connect the equipment as shown in the setup figure. Use a 50
n
8-12
Testing Performance
feedthrough termination 0.1 %.
2. Set up the HP 81 12A as follows:
Input Mode
Control Mode
Transition
PER
DEL
DTY
HIL
LOL
NORM
Off
Fixed
100 ms
6.5 ms
503
+0.10 v
0.00 v
3. Set up the DVM as follows:
Function
DCV
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......