HP-I B Verification
Test
Test Setup
•
Alter one or more other parameters and note the values.
•
Select R�L X (the number chosen in step 6).
•
Check that the recalled parameter set matches the standard
parameter set (RCLO) except for the change made to the standard
set in step 4.
System Controller
HP-IB HP B112A
o
Oscilloscope (HP 54121T)
Figure
8-22.
HP-IB Verification Test
Equipment
•
Controller (HP Series 200/300)
Procedure
Note
8-30
Testing Performance
•
Oscilloscope (HP 54121 T)
•
Cable Assembly BNC (2
x )
•
HP-IB cable
All program statements assume that the HP 81 12A is at HP-IB
address 12 and that BASIC 5.0/5.1 is being used.
1 . Connect the equipment as shown in the setup figure.
2. Use the following program statements to read the HP 81 12A
Standard Parameter Set:
DIM A$ [161]
REMOTE 712
CLEAR 712
OUTPUT 7 12 ; " CST"
ENTER 7 12 ; A$
PRINT A$
Allocate controller memory to receive HP
81 12A status string
Set HP 81 12A to remote mode
Clear HP 81 12A status and select stan
dard parameter set
Request current settings from HP 81 12A
Read the HP 81 12A settings
Display the HP 81 12A settings
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......