Double Pulse
Performance Test
Note
Specifications
The DBL measurement is described as the time interval between the
two leading edges of the double pulse, recorded at 50% amplitude.
Range
20 ns to 950 ms
Accuracy
±5% of programmed value ±2 ns
Repeatability
Factor 4 better than accuracy
Max Jitter
0.2% of programmed value ± lOOps
Equipment
•
Digitizing Oscilloscope with Accessory (HP 54121 T)
•
Counter (HP 5335A)
•
Cable, 50
n,
BNC to BNC, coaxial, 2 each (HP 8120-1839).
•
20 dB attenuators
•
son
feedthrough termination
(Required if counter input impedance f 50
n)
Test Setup 1
Oscilloscope (HP 5412xTI
20 dB Attenuator
HP 8112A
with BNC/SMA Adaptor
Figure
8-3.
Double Pulse Performance Test
8·6
Testing Performance
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......