Period Performance
Test
Specifications
Range
20 ns to 950 ms
Accuracy
±5% of programmed value ±2 ns
Repeatability
Factor 4 better than accuracy
Max Jitter
0.2% of programmed value ± lOOps
Equipment
•
Counter
(
HP 5335A)
Test Setup
•
Cable Assembly BNC
•
50 n Feedthrough Termination
(
Required if counter input impedance
=/;
50 n)
HP 8t12A
0
0
0
Counter
(H P 5 33 5A)
Figure
8·1.
Period Performance Test
Procedure
1 . Connect the equipment as shown in the setup figure. Use a
50 n feedthrough termination if you cannot select 50 n input
impedance on the counter.
2. Set up the HP 81 12A as follows:
Input Mode
Control Mode
Transition
DTY
HI1
101
NORM
Off
Fixed
50%
1 .00 v
- 1 .00 v
3 . Set the counter function control to PER A.
4. Set the HP 8 1 12A Period to the following values and read the
actual output from the counter. Record your results on a copy of
Testing Performance
8-3
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......