Note
9-B Adjustment Procedures
9. Centre the pulse base line vertically (offset = 8 V) and set
10. Set the VARIABLE LEVELS
=
-5 to +53
1 1 . Check overshoot < 53 in normal and complement modes.
Re-adjust A1C530 if necessary.
12. Set the HP 81 12A as follows:
Transition
Fixed
Pulse Width
_____
Overshoot
50%
t0%
0%
Pre-shoot
---,
Amplitude
50%
LOL
Figure 9-3. Typical Pulse Display
13. Repeat steps 7 to 1 1 using 90 - 1 103 / - 10 - 103 and adjust
A1C410 for overshoot < 103 in both normal and complement
modes.
14. Set the HP 8112A as follows:
LOL
LIMIT
-2.0
v
Off
15. Check the overshoot at both HIL and LOL for both Fixed and
Linear modes. If necessary, re-adjust A1C530 or A1C410 to
achieve the above limits. Repeat steps 1 to 12 if necessary.
The oscilloscope trace :flatness error ( GaAs input circuit) may affect
measurement of pre- and over-shoot
16. Set the HP 81 12A as follows:
Transition
HIL
LOL
LIMIT
Fixed
+0.5
v
-0.5
v
Off
17. Remove one of the two 20 dB attenuators and set attenuation
factor to 10.
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......