MODEL 8112A
ERRATA (Cont.)
page 8-12, Testing Performance
Output Levels Performance Test change to read:
step 3. Set up the DVM as follows
Function
DCV
Add or
Change:
Measurement
Peak Voltage
HP 3458A setup for peak-voltage measurement:
Function
DCV
MENU ->
Trig
EXT
(press Trig button, select EXT, terminate
with ENTER key)
MENU
->
NPLC
0.1
(press NPLC button, press period ".",
then " I " keys, terminate with ENTER key)
HP 3456A setup for peak-voltage measurement:
Function = V
TRIGGER EXT
N CYC INT
0. 1
(press period ". ", then " l " keys,
press STORE button,
press CHS (N CYC INT) key)
page 9-2, Adjustment Procedures
Table 9-1 , Adjustment Procedures - Changeable Components change to read:
Procedure
Reference
Range
Description
Overshoot/Transition Times Al C532
0 pF (open) to 33 pF
Increasing V532 dereases
transition times and
increases overshoot in
FIXED and LINEA mode
1 0. July 1 996
Page 7 of 1 6
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......