Slope
Equipment
•
Oscilloscopes (HP 54121 T /HP 54503A)
•
20 dB attenuators (3
x)
Procedure
1 . Set up the HP 8 1 12A as follows:
Trigger Mode
NORM
Control Mode
Off
PER
5 ms
DEL
65 ns
DTY
50%
Transition
Linear
LEE
100 µs
TRE
500 µs
HIL
+4.00
v
LOL
-4.00
v
COMPL
Off
DISABLE
Off
LIMIT
Off
STO
1
Oscilloscope (H P 54121T)
HP
8112A
Attenuators, 40d B (2x20)
20dB
Figure 9-7. Slope test setup
2. Set the HP 81 12A as follows:
LEE
STO
999 µs
2
3. Set the HP 8 1 12A as follows:
LEE
500 µs
STO
5
4. Set the HP 8 1 1 2A as follows:
TRE
100 µs
STO
3
5. Set the HP 8 1 12A as follows:
Adjustment Procedures 9-19
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......