Overshoot
&
Transition Time
Adjustment
Equipment
•
Oscilloscope (HP 54121T) .
•
Attenuator 20 dB (3
x
).
Procedure
1 . Set up the HP 81 12A as follows:
Trigger Mode
NORM
Control Mode
Off
PER
200 ns
DEL
65 ns
DTY
503
Transition
LEE
TRE
HIL
Linear
15.5 ns
15.5 ns
8 V
-8
v
Off
LOL
COMPL
DISABLE
LIMIT
Off (=Enable)
Off
2. Connect the HP 81 12A main output to the oscilloscope input 4,
via two 20 dB attenuators ( 40 dB) .
3. Connect the H P 8 112A trigger output t o the oscilloscope trigger
input via 20 dB attenuation.
4. Press
5. Adjust A1R535 to give best slope linearity in both COMPL off
(=Norm) and COMPL on modes.
6. Set the HP 81 12A as follows:
LEE
TRE
5.5 ns
5.5 ns
7. Set the oscilloscope as follows:
a. Full screen display (center one pulse horizontally and
vertically on the screen).
b. Number of averages to 64.
c. Attenuation factor to 100.
d. Select the
and turn the voltage markers on.
e. Set VARIABLE LEVELS = 95 - 1053 and press the Auto
Level Set.
f. Center the pulse top vertically (offset = 8 V).
g. Set V /div to 500 mV.
8. Adjust A1C530 for overshoot
<
53 in normal and complement
modes.
Adjustment Procedures 9-7
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......