Equipment
Test Setup 1
Test Se1up
2
Max Jitter
0.2% of programmed value ± lOOps
•
Counter
(
HP 5335A
)
•
Cable Assembly BNC
(
2
X)
•
Digitizing Oscilloscope
(
HP 54121T
)
•
Attenuator 20 dB, 2 W.
(
2 x )
•
50 !lFeedthrough Termination
(
Required if counter input impedance
-f;
50
fl).
Oscilloscope (HP
5412xTI
HP 8112A
20 dB Attenuator
with BNC/SMA Adaptor
Figure
8·5.
Pulse Width Performance Test
H P 8 1 1 2A
0
0
0
Counter
(H P 5 33 5A)
Figure
8-6.
Pulse Width Performance Test
Procedure
1 . Set up the HP 81 12A as follows:
Input Mode
NORM
Control Mode
Off
Transition
DEL
HIL
LOL
Fixed
65 ns
1 .00 v
- 1 .00 v
2. Connect the HP 81 12A and oscilloscope as shown in Figure 8-5.
3 . Verify that for the following HP81 12A WID settings the scope
display indicates that the HP 81 12A output is within the specified
limits.
4. Record your results on a copy of the Test Record, specified limits
are given here and on the Test Record.
Testing Performance
8-9
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......