Pre Adjustments
Equipment
•
Oscilloscope (HP 54121 T).
•
Attenuator 20 dB (3
x
).
Procedure
Oscilloscope (H P 54121TI
HP 8112A
0
Attenuators, 40dB (2x20)
20dB
Figure 9-2. Pre adjustments setup
Minimum Pulse Droop
1 . Set up the HP 81 12A as follows:
Trigger Mode
Control Mode
PER
DEL
DTY
Transition
LEE
TRE
HIL
LOL
COMPL
DISABLE
LIMIT
NORM
Off
1 ms
65 ns
503
Linear, Fixed, Gaussian as reqd.
10 ns
10 ns
+4.99
v
-4.99
v
Off
Off (
=
Enable)
Off
2. Connect the HP 81 12A main output to the oscilloscope input 4
via 40 dB (2
x
20 dB) attenuation.
3. Connect TRIGGER OUT of the HP 81 12A to the TRIG IN of
the oscilloscope, via a 20 dB Attenuator.
4. On the oscilloscope:
a. Press
and set for 2 V /division.
b. Set the attenuation factor to 100 and its offset to 0 V.
5. Adjust A1R515 for best pulse droop in all three transition
modes.
Adjustment Procedures 9-5
Summary of Contents for 8112A
Page 6: ......
Page 24: ......
Page 36: ......
Page 44: ......
Page 70: ......
Page 92: ......
Page 144: ...8 32 Testing Performance ...
Page 176: ......
Page 190: ......
Page 194: ...10 2 4 Servicing the Power Supply ...
Page 196: ......
Page 197: ......
Page 216: ......
Page 240: ...10 4 16 Servicing the Shaper and Output Amplifier ...
Page 242: ...A R l BD RY MR I N 1 2 7 I I IJt 51 CR1 B e i 1 _ _ _ _ 20 ...
Page 244: ......
Page 251: ......
Page 262: ...A B A2 ED RY CONTROL ...
Page 270: ......
Page 290: ...R B c R3 BD RY M I CROPROCESSOR 1 2 3 FIGURE 1 0 7 8 ...
Page 294: ... 8 X2l 2 X2 Figure A 1 Mechanical Parts 1 A 2 Replaceable Parts ...
Page 295: ...MP MP3 MP MP2 MP4 MPS MP 1 2 MPS MP MP I B ElElElEl Replaceable Parts A 3 ...
Page 334: ......
Page 378: ...Figure B 14 Mechanical Parts 1 B 44 Backdating ...
Page 380: ......
Page 382: ......
Page 402: ......