Statistical Calculations
The statistical calculation results are automatically cleared at the following times:
• When the user clears the statistical calculation results
• When the user clears the statistical calculation results
• When the user changes the measurement conditions (temperature compensation, scaling, NULL)
• When the user changes the comparator settings (p. 53)
• When the user changes the BIN settings (p. 57)
• When the instrument is reset (p. 93)
• When measurement conditions are loaded using the panel load function
• When the instrument is turned off (p. 28)
Printing
Touching
prints the statistical calculation results.
If there is no valid data, only the number of data points will be printed. If there is one valid data
point, the sample standard deviation and process capability index will not be printed.
If you do not wish to clear the statistical calculation results every time they are
printed
(Measurement screen)
MENU
>
I/F
1
2
3
ON
Automatically clears the statistical
calculation results every time they
are printed.
OFF
Does not clear the statistical
calculation results.
(Default setting)