39
Starting Measurement
Trigger function settings
Trigger source
You can set whether to enable trigger input from an external device. Setting the trigger source to
EXTERNAL
enables use of the EXT I/O TRIG terminal as well as the
*TRG
command. The default
setting is
INTERNAL
(
RUN
state).
Delay
You can set the delay time from trigger input until the start of measurement from 0 ms to 9999 ms
in increments of 1 ms. The default setting is
PRESET
(0 ms).
Adjust the delay time for measurement targets that require some time to respond. Set a long
delay time at first and then gradually shorten the time while viewing measured values.
Measurement count
You can set how many measurements to perform for each trigger event from 1 to 5000. The default
setting is 1. This setting is disabled while the instrument is in the
RUN
state.
(Measurement screen)
MENU
>
TRIG
>
SOURCE
1
1
2
Setting the trigger source
INTERNAL
Sets the instrument to the RUN
state (default setting).
EXTERNAL
Enables trigger input from an
external device.
2
1
2
3
Setting the delay
PRESET
No delay time (0 ms)
(default setting)
MANUAL
Sets the delay time.
˄
Increases the setting by 1.
˅
Decreases the setting by 1.
(Valid setting range: 0 ms to 9999 ms)
3
Measurement