Functional Specifications
Measured value format
Operation
Changes the format of responses to measured value queries.
RANGE FIX setting: Exponential part fixed based on the
measurement range
FLOAT setting: Floating point
(When using the FLOAT setting, the instrument automatically
transitions to the STOP state when transitioning to the
REMOTE state.)
Setting
Measured value format: RANGE FIX, FLOAT
Default setting
Measured value format: RANGE FIX
SCPI ID
Operation
Sets the response string for the
*IDN?
query.
Setting
SCPI ID: Up to 127 characters
Default setting
Blank (HIOKI, model name, serial number, software version)
Remote
When communicating with the LAN, USB, RS-232C, or GP-IB interface, places the
instrument in the remote state and disables touch panel and key operations.
The
[TRIG]
key functions except while the instrument is in the RUN state.
The remote state can be canceled as follows:
•By pressing the
LOCAL
key on the touch panel
•By cycling the instrument’s power
•By sending the :
SYSTem:LOCal
command with the LAN, USB, RS-232C, or GP-IB
interface
•By sending the
GTL
command with the GP-IB interface
Startup settings
Operation
Selects which settings to apply when the instrument is turned
on.
Settings
Startup setting: LAST, FACTORY, PANEL
Panel: No. 01 to No. 30
Default settings
Startup setting: LAST
Panel: No. 01
Reset
Reset
Reverts all settings other than panel data and interface
settings to their factory defaults.
(Operation is the same as for the
*RST
,
:SYSTem:PRESet
,
:STATus:PRESet
commands.)
System reset
Reverts all settings to their factory defaults.