Measurement Specifications
Parameter
Description
Time
T
6
Acquision time (RUN state)
50 Hz power 60 Hz power
FAST (1PLC)
26.9 ms
23.5 ms
MEDIUM (10PLC) 245 ms
205 ms
SLOW (100PLC)
3.92 s
3.37 s
Integration time other than above: Integration time + 5 ms
T
7
Calculation time
0.1 ms
T
8
From EOM signal output to TRIG signal
input
1 ms or greater
T
9
EOM pulse width (other than RUN state)
1 ms to 100 ms
T
10
EOM pulse width (RUN state)
50 Hz power frequency
T
I
=0.02PLC to 1PLC: 32.8 ms
T
I
=10PLC, 100PLC: 164 ms
T
I
=ms setting INT{(
T
I
+39)×0.025}×32.8
60 Hz power frequency
T
I
=0.02PLC to 1PLC: 29.4 ms
T
I
=10PLC, 100PLC: 147 ms
T
I
=ms setting INT{(
T
I
+39)×0.025}×29.4
T
I
: Integration time
INT (value): Rounds off the decimal portion of the value.