141
Timing Chart
(3) When the trigger source is set to
INTERNAL
while the instrument is in the
RUN
state
The EOM signal will generate pulse output (with an output time measured in milliseconds). Once
the HI, IN, LO, ERR, OB, or BIN0 to BIN9 signal turns on, it will remain on when measurement
starts and will continue in that state until the next measurement completes.
EOM
HI, IN, LO, ERR
OB, BIN0 – BIN9
ON
OFF
OFF
T10
T3
T4
T6
T7
Contact
check
Measurement
processing
Delay
Measurement Calculation
Self-
calibration
Measurement speed can be maximized with the following settings:
Setting
For more information
Contact check (
CONTACT CHECK
)
OFF
Trigger delay (
DELAY
)
0 ms
“Trigger measurement (measurement with user-
In this case, t10 = 0 ms.
Panel load timing
It is necessary to hold the LOAD signal for approximately 10 ms. The TRIG signal will be ignored while the
panel load function is being completed.
LOAD0 to LOAD4
State
Panel 1
Panel 1
Min. 10 ms
Max. 70 ms
(When switching to the 100 mV range, max. 700 ms)
Load processing
Panel 2
Panel 2
External Control (EXT
I/O)