54
Comparator Measurement (Using a Single Judgment Standard)
Judgment result
Set upper limit and
lower limit values
Statistics display
The judgment result and statistics sub-
display are shown on the Measurement
screen.
P1
,
P2
Toggles the statistics display.
CLR
Deletes the statistical calculation
results.
Setting the comparator function to ON automatically sets the BIN function to OFF.
The following judgments may be shown when the instrument is not able to perform measurement
normally:
Measured value display
Judgment
+OvrRng
HI
(if only a lower limit value was set,
IN
)
−OvrRng
If the absolute value judgment function is disabled:
LO
(if only an upper
limit value was set,
IN
)
If the absolute value judgment function is enabled:
HI
(if only a lower limit
value was set,
IN
)
NoCntct
or
−−−−−−−
−−
(No judgment)
See “3.5 Measurement Error Displays (Displays Other Than Measured Values)” (p. 46)
.
The Settings screen can be displayed from the Measurement screen’s sub-display.