![Hioki DM7275 Series Instruction Manual Download Page 161](http://html1.mh-extra.com/html/hioki/dm7275-series/dm7275-series_instruction-manual_2137275161.webp)
161
Functional Specifications
Key lock
Operation
When set from the front panel, all operations other than the
cancellation key are disabled.
Operation of the front panel is disabled while the KEY_LOCK
signal is being input and while a valid LOAD signal is being
input.
The
[TRIG]
key functions while the key lock is engaged.
Setting
ON, OFF
Default setting
OFF
Backlight
Setting
Brightness: 0% to 100% (in 10% steps)
Default setting
Brightness: 80%
Clock
Auto calendar, automatic detection of leap years, 24-hour format
Clock accuracy
±4 min/month
Default state
00:00, January 1, 2015
Supplied power supply
frequency
Settings
50 Hz, 60 Hz, AUTO
Default setting
AUTO
Output format
Setting
Date: YYYYMMDD, DDMMYYYY, MMDDYYYY
Date delimiter: Slash, hyphen, period
Decimal point: Period, comma
Data delimiter: Comma, semicolon, tab, space
(Setting applies to screen display, USB flash drive output,
USB keyboard output, and printer output.)
Default settings
Date: YYYYMMDD
Date delimiter: Slash
Decimal point: Period
Data delimiter: Comma
Self-test
ROM test, RAM test
Buzzer
Settings
Volume: OFF, SMALL, MEDIUM, LARGE
Key tone: ON, OFF
Auto-hold tone: ON, OFF
Error tone: ON, OFF
Default settings
Volume: MEDIUM
Key tone: ON
Auto-hold tone: ON
Error tone: ON
Touch panel adjustment
Adjusts for touch panel misalignment by setting the location of the top left and bottom
right corners of the touch panel.
The settings can be reverted to their factory defaults.
Measurement
information
Displayed information
Displays instrument settings.
Communications
monitor
Operation
•Displays the data being sent and received with the LAN,
USB, RS-232C, and GP-IB interface.
•Saves sent and received commands on the USB flash drive
(log function).
Setting
Communications monitor: ON, OFF
Log: ON, OFF
Default setting
Communications monitor: OFF
Log: OFF
13
Specifications
www.
.com
1.800.561.8187