Measuring Multiple Targets
Taking steps to prevent short-circuits
Exercise caution concerning the following so as to avoid shorting the measurement target:
(1) Design the switching device so that all contacts are off when it is turned on and off.
(2)
Allow an interval of time when all contacts are off when switching contacts (“break before
make”).
(3) Insert a fuse into the measurement line.
Avoid use of fuses with a rating of 1 A or less and resettable fuses as they have a large thermal
electromotive force.
RY1
RY2
RY3
HIGH
LOW
V
Instrument
10
9
8
7
6
5
4
3
2
1