68
Obtaining Stable Measured Values
1
1
2
2
Select the unit.
3
(When
PLC
is selected)
Select the integration time.
(100 PLC, 10 PLC, 1 PLC, 0.2 PLC, 0.02 PLC)
(When
ms
is selected)
Select the integration time.
˄
Increases the value by 1.
˅
Decreases the value by 1.
(Valid setting range: 1 ms to 9999 ms)
•If the instrument is susceptible to the effects of the external environment: See “Appx. 4 Noise
•When the integration time is set to 0.02 PLC, an integration time of 0.4 ms will be used
regardless of the power supply frequency.
•Even when a longer integration time is used, fluctuations on the order of several microvolts may
be observed due to fluctuations in thermal electromotive force and the effects of burst noise.
See “Appx. 3 Causes of Error in Voltage Measurement” (p. Appx.5)